Section 6 -
Page 3 of 4
3121-9001-0167_EN.03.00_QT_Manual
04/2016
MESSAGE OR
PROBLEM
DESCRIPTION
RESOLUTION
Invalid Result
(2)/(1)
(Error 107)
A result was not able to be
calculated
Remove the cartridge and repeat with a
new cartridge.
107 errors are most commonly seen when
a cold cartridge is used with the system
Make sure the cartridge has sufficient time
to reach the same temperature as the
analyzer before use
107 errors can sometimes occur when the
analyzer is working at the bottom of it’s
temperature range and the blood volume/
hemoglobin concentration is towards the
bottom of the acceptable range. Make
sure that the temperature is within the
working range of the analyzer and that the
collected sample fills the slot on the blood
collector*
Invalid Barcode Format
(Error 108)
The calibration barcode has not
been recognized
Calibration Barcode:
re-scan the calibration
barcode or scan a barcode from a different
carton.
Cartridge Lot Barcode:
open and close the
analyzer door. If the error is still displayed, use
a different cartridge*
High Background
Fluorescence
(Error 110)
The test cartridge is faulty
Remove cartridge and repeat test with a
new cartridge*
Curve Fit Fail
(Error 112)
A result was not able to be
calculated
Invalid Result (2)
(Error 120, 121,
122, 123,124,
125 or 126)
A result was not able
to be calculated
Mix Motor Stall
(Error 128)
Mixer Motor has failed to start
during the assay
Remove cartridge, close the door and allow
the system to return to the home screen.
Repeat the test with a new cartridge*
*
If the problem continues, contact your local distributor or customer support.