DK50
DK50
PLUS
NP-DK50-3_07-2014-md
- 30 -
7/2014
Dental compressor with base DK50 PLUS/M
(Fig.6)
After removing all packaging material, place the product on the floor and remove stabilization parts X and Y
(Detail A). Direct the output pressure hose and power cord out the back of the compressor. Connect the
condensate drain hose to the vessel (20). Install the tank into the bracket on the compressor.
Dental compressor in box DK50 PLUS S
(Fig.4, Fig.6)
After removing all packaging material, place the product on the floor and remove stabilization parts X and Y
(Detail A). Place the wall-mounted stopper (27) - 2 pcs onto the compressor housing in the rear top part of
the housing and put the housing onto a required place. The stoppers provide a sufficient distance of the
housing from a wall for thorough ventilation. For setting up the compressor in housing you must open the
door on the housing using the attached key and remove connecting reinforcement (26) in the front bottom
part of housing. If necessary, the door may be disassembled using door hinge (31). Protrude pressure hose
via a hole in housing and connect it to the appliance in a suitable way. Grasp the compressor at its handle
and put it into the casing using built-in wheels (32). Fix the hose of a manometer in a hose into the fast-on
coupling on a compressor, put the connecting reinforcement (26) back and connect the pressure hose to a
compressor. Insert the electric power cord of a compressor into a socket (33) on a housing. By slight rotation
of rectification screws (34) set the correct position of door against the casing frame. When closing the door
the pin (23) on the door must easily snap in the opening in the casing frame. Close the housing doors and
duly lock the lock (25). Connect the mains plug into the mains socket.
It is not allowed to leave the key in a lock! It must be saved against non-instructed persons!
Dental compressor in box DK50 PLUS S/M
(Fig.4, Fig.6)
After removing all packaging material, place the product on the floor and remove stabilization parts X and Y
(Detail A). Place the compressor into housing similarly as in the previous paragraph. Prior to placing the
compressor into housing, protrude house for condensate drain via hole in housing and connect it to a bottle
(20). The magnetic holder (30) with a vessel (20), for entrapping condensate from a dryer may be fixed at the
sides of housing or from the front on its doors. When fixing the holder with a vessel at the housing side it is
necessary to consider a space of at least 11 cm between the housing and furniture. Distance smaller than
the specified one may cause problem with handling of the vessel.
The vessel must always be installed so that the lower section is near the floor; any other
installation may damage the dryer!
9.2.
Compressed air outlet
(Fig.7)
At the compressed air outlet (1) of the compressor, connect the pressure
hose with a nut (2) and fasten clip (3). Connect the hose to the dental
appliance.
9.3.
Electrical connection
Plug the electrical cord into the mains.
The appliance is equipped with a grounded plug. Make sure this connection complies with
local electrical codes. The mains voltage and frequency must comply with the data stated
on the appliance label.
Keep the socket easily accessible to ensure that in an emergency the appliance can be safely
disconnected from the mains.
Connection to the power distribution box must be max.16 A.
Electrical cable may not contact the hot parts of a compressor. Insulation could be
damaged!
If any electrical cord or air hose is damaged it must be replaced immediately.
Fig.7