PRODUCT MAINTENANCE
02/2021
25
NP-DK50 B-A-1_02-2021
PROCESS:
Jumper removal (B-A process)
Disconnect the product from the mains by pulling the plug out of the socket.
Remove the cover on the electrical panel (on the compressor).
The jumper is in the terminal strip
– B
Remove the jumper from the terminal strip
– A
Reinstall the cover on the electrical panel.
Insert the compressor into the cabinet.
Connect the cabinet to the compressor by inserting the power cord with connector into a power
socket.
Connect the device to the electrical mains
Activate the compressor by turning on the switch on the pressure switch and the switch (4) on
the cabinet (Fig. 5)
Compressor with cabinet - when operating in the cabinet, there
must not be
an electrical
connection. (Fig. 9 pos A).
Note:
In case of using a compressor without a cabinet or in case of maintenance, a jumper
which is not installed into the terminal strip should be kept
in a designated place next to the
electrical panel.
16.11. Cleaning and disinfection of the exterior surfaces of the product
Clean and disinfect the exterior surfaces with neutral cleaning products.
Use of aggressive detergents and disinfectants containing alcohol and chlorides
can lead to surface damage and discolouration.
17. LONG-TERM SHUTDOWN
If the compressor will not be used for a prolonged time period, drain any condensate from the air
tank. Then turn on the compressor for 10 minutes, keeping the drain valve open (1) (Fig. 6). Switch
off the compressor by switch (2) at pressure switch (1) (Fig. 5), close the drain valve and disconnect
the appliance from the mains.
18. DISPOSAL OF DEVICE
Disconnect the equipment from the mains.
Release the air pressure in the pressure tank by opening the drain valve (1) (Fig. 6).
The components of the product are non-toxic.
Dispose of the appliance following all environmental regulations.
Summary of Contents for DK50 B
Page 1: ...DK50 B User manual Instrukcja obs ugi N vod na pou itie...
Page 2: ......
Page 4: ...CONTENTS 4 28 SPIS TRE CI 51 OBSAH 74...
Page 30: ...NP DK50 B A 1_02 2021 30 02 2021 3 3 1 3 2...
Page 31: ...02 2021 31 NP DK50 B A 1_02 2021 4 4 1 4 2 4 3 4 4...
Page 32: ...NP DK50 B A 1_02 2021 32 02 2021 5 5 1 25 55 C 24 70 C 90...
Page 34: ...NP DK50 B A 1_02 2021 34 02 2021 1 DK50 B 1 A a 2 3 4 5 6 7 8 9...
Page 36: ...NP DK50 B A 1_02 2021 36 02 2021 8...
Page 38: ...NP DK50 B A 1_02 2021 38 02 2021 10 10 1 1 2 4 11 16 A 11 1 11 2 2 1...
Page 40: ...NP DK50 B A 1_02 2021 40 02 2021 40 C 32 C...
Page 44: ...NP DK50 B A 1_02 2021 44 02 2021 13 16 2 a e a 40 C 16 3 B2 40 C EV1 EV2 40 C a 16 4...
Page 45: ...02 2021 45 NP DK50 B A 1_02 2021 16 5 1 1 6 DK50 BS 16 6 Sicherheitsventil berpr fen 2 1 2 7...
Page 47: ...02 2021 47 NP DK50 B A 1_02 2021 9 B 9 230 V 16 10 9...
Page 48: ...NP DK50 B A 1_02 2021 48 02 2021 9 11 1 11 2 B A B A 4 5 9 A 16 11 17 10 1 6 2 1 5 18 6...
Page 49: ...02 2021 49 NP DK50 B A 1_02 2021 17 10...
Page 50: ...NP DK50 B A 1_02 2021 50 02 2021 a e 19...
Page 101: ......
Page 102: ......
Page 103: ......