3 OPERATING
3.1 TECHNICAL FEATURES
The specifications below are given after at least 30 minutes use within the specified operating
temperature range.
Operating
Constant voltage
Automatic
Constant current
Automatic
Mini maxi adjustment
Voltage
0 to 32.00 Volts (0 to ±10mV)
Current
0 to 6.000 Amps
OVP (voltage protection)
0 to 32.20 Volts
OCP (current protection)
0 to 6.100 Amps
Adjustment accuracy
± (% of offset)
Voltage
< 0,03% +10 mV
Current
< 0,03% +2 mA
Regulation / Load 10
–
90%
Constant voltage
< 20 mV
Constant current
< 1 mA
Regulation / Source ±10%
Constant voltage
< 1 mV
Constant current
< 1 mA
Ripple
Constant voltage
< 1.5 mV
RMS
; < 5mVp-p noise
< 10 mVp-p Pics of commutation
Constant current
< 0.4 mA
RMS
ou 1mAp-p
Accuracy measurement (25°C ±5°C)
± (% of offset)
Voltage
< 0,06% or ±10 mV
Current
< 0,06% or ±10 mA
Temperature coefficient
± (% of offset)
Voltage
0,01% / °C
Current
0,05% / °C
Resolution
Voltage / Current
4 digits
Time of answer (Load variation)
Load 50
–
100%
< 8 ms (±20mV)
Load 100
–
50%
< 5 ms (±20mV)
Overvoltage output
ON/OFF source or output
< 0.3V
Voltage programming speed (up) to 1% of the total course
Without load
Load 100%
Rise times
0
–
32 V
70 ms
420 ms
0
–
16 V
35 ms
128 ms
Fall times
32V
–
0V
100 ms
8.5 ms
16V
–
0V
70 ms
7.5 ms