background image

ELCO S.r.l. - ELK 38 - OPERATING INSTRUCTIONS - Vr. 02 - ISTR 06261 - PAG. 9

 

4.11 - LOOP BREAK ALARM FUNCTION 
All  the  parameters  referring  to  the  Loop  Break  alarm  function  are 
contained in the group 

]

LbA”. 

The  Loop  Break  alarm  is  available  on  all  the  instruments,  which 
intervenes  when,  for  any  reason  (short-circuit  of  a  thermocouple, 
thermocouple  inversion,  load  interruption),  the  loop  control  is 
interrupted. 
First of all, it is necessary to establish to which output the alarm has 
to correspond.   
To do this it is necessary to set the parameter relative to the output 
to  be  used  (“O1F”  ,  “O2F”)  in  the  group  

]

Out”,  programming  the 

parameter as : 
=  ALno  if  the  alarm  output  has  to  be  ON  when  the  alarm  is  active 
while it is OFF when the alarm is not active. 
=  ALnc  if  the  alarm  output  has  to  be  ON  when  the  alarm  is  not 
active while it is OFF when the alarm is active. 
=  ALni  if  the  alarm  output  has  to  be  ON  when  the  alarm  is  not 
active, while it is OFF when the alarm is active but with reverse led 
indication (led ON= alarm OFF). 
Enter group  

]

LbA” and program which output the alarm signal has 

to be addressed to on par. “OLbA”,.  
The  Loop  Break  alarm  is  activated  if  the  output  power  remains  at 
the  100%  of  the  value  for  the  time  programmed  on  par.  "LbAt" 
(expressed in sec.). 
To  avoid  false  alarms,  the  value  of  this  parameter  has  to  be  set 
considering the time the plant takes to reach the Set point when the 
measured  value  is  a  long  distance from it (for example at the plant 
start-up).  
On  alarm  intervention,  the  instrument  visualizes  the  message 
“LbA” and behaves as in the case of a measurement error giving a 
power output as programmed on par.  “OPE” (programmable in the 
group 

]

InP”)

To  restore  normal  functioning  after  the  alarm,  select  the  control 
mode “OFF” and then re-program the automatic control (“rEG”) after 
checking the correct functioning of probe and actuator. 
To exclude the Loop Break alarm, set “OLbA” = OFF. 
 
4.12 - FUNCTIONING OF KEY “U” 
The function of key “U” can be set through par. “USrb” contained in 
the group “

]

PAn”. 

The parameter can be programmed as : 
= noF : no function 
=  tunE  :  Pushing  the  key  for  1  sec.  at  least,  it  is  possible  to 
activate/deactivate Auto-tuning or Self-tuning 
= OPLO : Pushing the key for 1 sec. at least, it is possible to swap 
from automatic control (rEG) to manual one (OPLO) and vice versa. 
=  Aac  :  Pushing  the  key  for  1  sec.  at  least,  it  is  possible  to 
acknowledge the alarm. (see par. 4.10.1) 
=  ASi  :  Pushing  the  key  for  1  sec.  at  least,  it  is  possible  to 
acknowledge an active alarm (see par. 4.10.1) 
 = CHSP : Pushing the key for 1 sec. at least, it is possible to select 
one of the 4 pre-programmed Set Points on rotation.  
=  OFF  :  Pushing  the  key  for  1  sec.  at  least,  it  is  possible  to  swap 
from automatic control (rEG) to OFF control (OFF) and vice versa. 
 
4.13 - PARAMETERS CONFIGURATION BY “KEY01” 
The instrument is equipped with a connector that allows the transfer 
from  and  toward  the  instrument  of  the  functioning  parameters 
through the device ELCO KEY01 with 5 poles connector. 
This  device  it’s  mainly  useable  for  the  serial  programming  of  the 
instruments which need to have the same parameters configuration 
or to keep a copy of the programming of an instrument and allow its 
rapid retransmission. 
To  use  the  device  KEY01  it’s  necessary  that  the  device  or 
instrument are being supplied. 
Instrument supplied and device not supplied 

 

Instrument supplied from the device 

 

To  transfer  the  configuration  of  an  instrument  into  the  device 
(UPLOAD) it is necessary to proceed in the following way:   
1) position both dip switch of KEY 01 in the OFF mode.   
2)  connect  the  device  to  the  instrument  ELK  plugging  the  special 
connector.   
3) verify that the instrument or the device are supplied  
4)  observe  the  indication  led  on  the  device  KEY  01:  if  it  results 
green  this  means  that  a  configuration  is  already  loaded  on  the 
device  while  if  it  results  green  blinking  or  red  blinking  this  means 
that it has not been loaded any valid configuration on the device .   
5) press the button placed on the device.   
6)  observe  the  indication  led  :  after  having  pressed  the  button,  the 
led  becomes  red  and  therefore,  at  the  end  of  the  data  transfer,  it 
becomes green.   
7) now it is possible to disconnect the device. 
To  transfer  the  configuration  loaded  on  the  device  onto  an 
instrument  of  the  same  family  (DOWNLOAD),  it  is  necessary  to 
proceed in the following way:   
1) position both dip switch of KEY 01 in the ON mode.   
2)  connect  the  device  to  an  instrument  ELK  having  the  same 
features  of  the  one  from  which  has  been  downloaded  the  desired 
configuration,   plugging the special connector.   
3) verify that the instrument or the device are supplied  
4)  observe  the  indication  led on the device KEY 01: it has to result 
green, because if the led results green blinking or red blinking, this 
means  that  on  the  device  it  has  not  been  downloaded  any  valid 
configuration and therefore it’s useless to continue. 
5) if the les results green, press the button placed on the device.   
6)  observe  the  indication  led  :  after  having  pressed  the  button,  the 
led  becomes  red  and  therefore,  at  the  end  of  the  data  transfer,  it 
becomes green.   
7) now it is possible to disconnect the device. 
For  additional  info,  please  have  a  look  at  the  KEY01  instruction 
manual. 
 

5 - PROGRAMMABLE PARAMETERS 

 
Here  following  are  described  all  the  parameters  available  on  the 
instrument. Some of them could be not present or because they are 
depending  on  the  type  of  instrument  or  because  they  are 
automatically disabled as unnecessary. 
 
5.1 - PARAMETERS TABLE
 
Group “SP” (parameters relative to the Set Point) 

Par.

 

Description

 

Range

 

Def.

 

Note

 

1

 

nSP

 

Number of the 
programmable Set 
point 

 

 1 ÷ 4

 

1

 

 

2

 

SPAt

 

Active Set point

 

1 ÷ nSP

 

1

 

 

SP1  Set Point 1

 

SPLL ÷ SPHL

 

0

 

 

SP2  Set Point 2

 

SPLL ÷ SPHL

 

0

 

 

Summary of Contents for ELK 38

Page 1: ...ACTION PID CONTROL 4 6 DOUBLE ACTION PID CONTROL 4 7 AUTO TUNING AND SELF TUNING FUNCTIONS 4 8 REACHING OF SET POINT AT CONTROLLED SPEED AND AUTOMATIC COMMUTATION BETWEEN TWO SET POINTS 4 9 SOFT STAR...

Page 2: ...er two seconds in the same condition the changing speed increases in order to allow the desired value to be reached rapidly Once the desired value has been reached by pushing key P it is possible to e...

Page 3: ...of the controller During automatic control it is possible to visualize the control power on the display by pushing key UP The range of the power values goes from H100 100 of the output power with rev...

Page 4: ...nearization J Ir J or K Ir CA for thermoresistances Pt100 IEC Pt1 or thermistors PTC KTY81 121 Ptc or NTC 103AT 2 ntc for normalised signals in current 0 20 mA 0 20 or 4 20 mA 4 20 for normalised sign...

Page 5: ...value SP HSEt Vice versa in case of direct action or cooling Func CooL it deactivates the output when the process value reaches SP HSEt in case of symmetrical hysteresis or SP in case of asymmetrical...

Page 6: ...control monitoring and the continuous calculation of the parameters during control Both functions automatically calculate the following parameters Pb Proportional Band tcr1 Cycle time of the output 1r...

Page 7: ...on only works through PID control and allows the limitation of control power when the instrument is switched on for a programmable period of time This is useful when the actuator driven by the instrum...

Page 8: ...ument is in alarm condition the alarm is not activated It will be activated only when the process value is in non alarm conditions and then back in alarm conditions ALARM DELAY the alarm output may be...

Page 9: ...CONFIGURATION BY KEY01 The instrument is equipped with a connector that allows the transfer from and toward the instrument of the functioning parameters through the device ELCO KEY01 with 5 poles conn...

Page 10: ...arm AL1L 9999 9999 27 HAL1 Alarm AL1 hysteresis OFF 9999 1 28 AL1d Activation delay of alarm AL1 OFF 9999 sec OFF 29 AL1i Alarm AL1 activation in case of measuring error no yES no Group LbA parameters...

Page 11: ...defines the error conditions of the input allowing the instrument to give the power programmed on par OPE as output The possibilities are Or the condition occurs in case of over range or probe breakag...

Page 12: ...with Func HEAt than SP SP 2 or higher with Func CooL than SP SP 2 OFF Autotuning disabled When it s occurring an Autotuning cycle led AT blinks SELF SELF TUNING ENABLE Parameter used to enable yES or...

Page 13: ...y to repeat the auto tuning LbA Loop control interrupted Loop break alarm Check the working of probe and actuator and swap the instrument to rEG control ErEP Possible anomaly of the EEPROM memory Push...

Page 14: ...C 58 0 230 0 F 0 20 mA SEnS 0 20 1999 9999 199 9 999 9 19 99 99 99 1 999 9 999 4 20 mA SEnS 4 20 1999 9999 199 9 999 9 19 99 99 99 1 999 9 999 0 50 mV SEnS 0 50 1999 9999 199 9 999 9 19 99 99 99 1 99...

Reviews: