R
16
The tile should be cleaned and the measurement of the calibration tile
repeated.
If scratches or permanent marks are seen, press the C key. The
calibration value will not be stored. A new calibration tile will be required if permanent marks are present.
7 SYSTEM SET-UP
Use system set-up to configure your gauge to suit the way you work.
To enter system set-up:
Press and hold the DOWN button. While continuing to press the DOWN
button, press and hold the C button until the display shows:
Now use the SCROLL UP/DOWN key to scroll through the set-up options
- see paragraphs 7.1 to 7.5.
7.1 CALIBRATION VALUE - CHANGING
The gloss value(s) of the calibration tile supplied with your gauge are stored within the gauge at the time of
manufacture. To see these value(s), look for the sticker inside the calibration tile case.
Normally these value(s) do not need to be changed, however if it becomes necessary to modify the stored
value(s), for example if the calibration tile is re-calibrated, or a different calibration tile is used, use the
following procedure:
ARE YOU SURE?
Press Read Key
System Set-up