R
28
Taking test readings
Press
YES
softkey (see previous section) to take
test readings. This allows the calibration of the
gauge to be tested without contributing to the
statistical calculations.
Figure 30. TEST READINGS screen
CAL
softkey returns the gauge to
Step 1
of the
calibration adjustment procedure.
OK
softkey exits the calibration adjustment
procedure and returns the gauge to the reading
screen.
Other calibration methods
For the 2-POINT, ROUGH and SPECIAL
SUBSTRATE calibration methods
Step 2
requires
readings to be taken on a thin standard value
instead of on an uncoated base.
Figure 31. Step 2 - On thin standard
Repeating the reading will display the average. This
is particularly useful for rough surfaces as it allows
variations in the surface to be accounted for in the
calibration adjustment, therefore improving the
accuracy of the gauge.
Figure 32. Step 2 - Calibration adjustment on
thin standard
First reading
Second reading