R
33
7 PROBES
An extensive range of probes is available for the
Elcometer 456
3
Coating Thickness Gauge. Probes
for ferrous (F), non-ferrous (N) and dual
ferrous/non-ferrous (FNF) operation are available
as either integral (built-in) or separate options.
Separate probes are fully interchangeable and are
available in standard, PINIP™ and miniature
formats.
PINIP™ format (Plug-In Integral Probe) is an
integral style probe which plugs in to a separate
gauge. This provides all the benefits of an integral
gauge and the flexibility of a separate gauge in a
single unit.
Miniature separate probes permit measurements
where space is restricted.
7.1 FERROUS (F) PROBES
F
probes measure the thickness of non-magnetic
coatings on magnetic substrates. They can be used
on paint, plastic, galvanising, enamel, powder paint,
hard chrome and other coatings such as electro-
less nickel applied to steel or iron.
7.2 NON-FERROUS (N) PROBES
K
N
probes measure the thickness of non-metallic
coatings on non-magnetic metals. They can be
used on anodising, paint, plastic coatings, powder
paint, etc. applied to aluminium, brass, non-
magnetic stainless steel, etc.
7.3 DUAL FERROUS/NON-FERROUS (FNF)
PROBES
FNF
probes are dual function, F and N in one probe.
FNF gauges will automatically detect the type of
substrate and set the mode accordingly.
Alternatively the mode can be set manually - see
“PROBE:” on page 20 and “Coatings on galvanised
or metallised steel” on page 35.
j.
When readings are combined a
symbol will be
displayed on the Reading Screen (Figure 35).
k.
Using an N probe (or an FNF probe manually set to
N1) on a ferrous substrate will give a reading, but the
reading will be incorrect.
Op_456_3_English.book Page 33 Thursday, January 22, 2009 7:58 PM