background image

7

R

5 GLOSS MEASUREMENT

5.1 General measurement tips

It is important that the test area is free of dust, dirt, fingerprints, etc.,
as these will affect the gloss.
Glossmeters integrate all the light going to the receptor so if there is
poor covering of the aperture and ambient light is able to get in, this
is also detected and gives anomalous high readings. When
measuring translucent materials, such as plastic or glass, transmitted
ambient light must be excluded by either working in a dark room or by
shielding with opaque cloth.
The optics in the Glossmeter are focused to the plane of the gauge,
so the test surface must be in contact with the base to obtain accurate
readings. This applies also to silver-backed mirrors, where the gauge
will be separated from the primary reflecting surface by the thickness
of the glass.
When measuring textured materials such as leather, gloss readings
may be lower than expected after a visual examination. This is
because the measurement is an integral of the highlighted areas and
the apparently dark ‘valleys’. For similar reasons, gloss on very
uneven surfaces will be unreliable.
Care must also be taken when testing curved surfaces. The 85º and
75º measurement angles are totally unsuited to this purpose and
should only be used on flat surfaces. The 60º, 45º, 45º/0º and 20º
measurement angles get progressively better at this task but will still
only give reliable readings on curved surfaces of >125 mm (5")
radius. Best results are obtained on cylinders when measuring
parallel to the axis. For accurate gloss measurements on curved
surfaces with a radius as low as 12 mm (½") a special gauge, the
Elcometer 400 NOVO-CURVE, is available - see “Related
equipment” on page 14.

Op_401_English.fm  Page 7  Tuesday, August 30, 2005  8:54 AM

Summary of Contents for Novo-Gloss 401

Page 1: ...English Elcometer 401 Novo Gloss Glossmeter Operating Instructions Op_401_English fm Page 1 Tuesday August 30 2005 8 54 AM...

Page 2: ...ents Ltd NOVO GLOSS is a trademark of Rhopoint Instruments Ltd Copyright Elcometer Instruments Ltd 2005 All other trademarks acknowledged All rights reserved No part of this Document may be reproduced...

Page 3: ...4 Calibration adjustment 5 4 1 The calibration tile 5 4 2 To calibrate your gauge 6 5 Gloss measurement 7 5 1 General measurement tips 7 5 2 Gloss measurement 8 6 Maintenance 8 6 1 Calibration tile 8...

Page 4: ...d held gauge for measuring the gloss on flat surfaces from matt to mirror finishes The Glossmeter is available in three versions single angle dual angle and triple angle These instructions apply to al...

Page 5: ...e ensure that this packaging is disposed of in an environmentally sensitive manner Consult your local Environmental Authority for further guidance Note Before you use your Elcometer 401 for the first...

Page 6: ...lows the battery to be charged The battery will be fully charged in approximately 16 hours from full discharge Ensure that the LO BAT symbol is showing on the display before recharging this will maxim...

Page 7: ...onds after the button is released The gauge then switches off to conserve battery power The last reading is lost when the gauge switches off 4 CALIBRATION ADJUSTMENT Before taking readings adjust the...

Page 8: ...3 Wait until the reading stabilises and then adjust the calibration thumb wheel until the reading on the display matches the value on the calibration tile corresponding to the chosen angle Note that...

Page 9: ...eflecting surface by the thickness of the glass When measuring textured materials such as leather gloss readings may be lower than expected after a visual examination This is because the measurement i...

Page 10: ...the Glossmeter and the certified calibration tile are a requirement of quality management procedures e g ISO 9000 and other similar standards With the exception of lamp replacement the gauge does not...

Page 11: ...ith dual or triple angle gauges Calibration tiles can be re calibrated by returning the tile to Elcometer your local Elcometer supplier or to a recognised gloss calibration facility such as BAM German...

Page 12: ...Figure 3 The arrows indicate the filament direction Replace the clamping ring and connect the new lamp Referring to Figure 4 centralise the calibration thumb wheel potentiometer A Take a reading on a...

Page 13: ...orientation shown in Figure 5 The arrows indicate the filament direction Replace the clamping ring and connect the new lamp Referring to Figure 6 centralise the calibration thumb wheel potentiometer A...

Page 14: ...ientation shown in Figure 7 The arrows indicate the filament direction Replace the clamping ring and connect the new lamp Referring to Figure 8 centralise the calibration thumb wheel potentiometer A T...

Page 15: ...measurements however over the life of the gauge replacements may be required The following replacement items are available from Elcometer or your local supplier Description Sales Part No 20 Gloss Sta...

Page 16: ...atic switch off delay 15 seconds Lamp life 200 hours nominal 10 RELATED EQUIPMENT In addition to the Elcometer 401 Novo Gloss Glossmeter Elcometer produces a wide range of other equipment for measurin...

Reviews: