1. PRESENTATION

VIP CLAMP MINI 

emphasises reliability and simplicity of use to respond to the needs of power professionals:

- A compact unit integrating the current sensor for intensity measurements without breaking the test circuit

- Outstanding ergonomic features, in particular: -automatic or manual selection of the type of signal to     be measured,

AC or DC,

  - measurement of the RMS value of any signal (AC+DC),

  - automatic selection of measurement calibre,

  - programmable audio voltage indication: V-Live,

  -“measurement range exceeded” indication,

  - Backlighting of the digital display

  - auto power off system, -MIN - MAX – PEAK value recording function,

  - correction of deviations in DC measurement (DC zero)

  - automatic compensation of measurement cable resistance (

W

 zero).

- Compliance with CEI electrical safety standards and CE markings

- Lightness and ruggedness for field use With all new additional features:

- “Inrush” function in particular for measurement of motor starting currents.

- Phase order indication function using “2-wire” technique PFISTERER Licence - (instead of 3-wire) capable of

  determination through contact only, with no fastidious connections.

2. DESCRIPTION

(vedi schema § 8 Allegato)

     Jaws

6-way rotary switch:

OFF

      Deactivation of the clamp, activation is ensured by selection of other functions

          Measurement of DC and AC voltages (rms value)

             Continuity measurement, by use of the yellow resistance key and semi-conductor test
A           Measurement of DC and AC voltages (rms value)

PF 

        Measurement of active, reactive and apparent

var W~ 

powers and power factor in single phase or

VA

       balanced 3 phases systems (by use of the yellow key)

  

  

Selection of phase order indicator for 3- phase system with or without neutral

Command keys

The keys are capable of 3 types of action:

Short pressure

>1.3 s, it is valid if the key pressure is detected

Long pressure

> 1.3 s, this gives access to a measurement or operating mode. Holding or releasing the key has

 

no effect.

Held pressure

Gives access to a measurement or operating mode and remains in this mode as long as pressure is held. Releasing

the key causes return to the previous mode.

· HOLD

  has 4 different functions (see description § 3.12):

   - Display lock

   - Preselection of MIN/MAX mode

   - Automatic compensation for cable resistance

   - Automatic compensation of current measurement zero

· The yellow key

 has 6 different functions (see also description § 3.12):

  

 - Manual selection of AC, DC or AC+DC mode

   - Selection of INRUSH function -Selection of resistance function (

W

), semi-conductor test     ontinuity test ( )

   - Measurement of active, reactive, apparent powers and power factor calculation -Single phase or balanced three

     phases systems selection

   - Phase order indication measurement (see description § 3.11)

·

 

MIN/MAX

 operates by end-around shift on short pressure:

MIN/MAX

Funzioni  V e A

Altre

Funzioni

PEAK value

1a Press

2 a Press

3 a Press

4a Press

MAX value

MIN value

Return to PEAK value

MAX value

MIN value

Return MAX value

_

1

Summary of Contents for VIP CLAMP MINI Series

Page 1: ...e clamp box is not properly closed Before any measurement check the proper positioning of the cables and switch When measuring current check for proper alignement of the conductor in relation to the m...

Page 2: ...ement zer 9 3 12 5 Manual selection of AC DC or AC DC mode 9 3 12 6 Selections possible in continuity function 9 3 12 7 Selectione OF INRUSH function 9 3 12 8 Suppression of automatic stop 9 3 12 9 Ac...

Page 3: ...ty measurement by use of the yellow resistance key and semi conductor test A Measurement of DC and AC voltages rms value PF Measurement of active reactive and apparent var W powers and power factor in...

Page 4: ...tal display The liquid crystal display includes the digital display of the measured values the related units and symbols Digital display 4 digits 9999 points 3 decimal points and signs DC and peak mea...

Page 5: ...V 3 Connect the assembly to the voltage source to be measured making sure if possible that this voltage does not exceed the maximum acceptable limits see table below Range switching and AC DC selecti...

Page 6: ...asuring range Accuracy 1 Resolution Open circuit voltage Protection 0 0 399 9W 1 L 2 pto 0 1 W 300 A Measuring current 3 2 V 500V AC o 750V DC or peak MIN MAX Mode Accuracy same as previous table 0 2...

Page 7: ...o in DC Repeatability of the measurement after several consecutive closings of the clamp 0 3 typical MIN MAX Mode Accuracy same as previous table 0 2 L Capture time 100 ms typ PEAK Mode Accuracy same...

Page 8: ...3f arrow on the jaws must be directed in the direction of power circulation from the source to the load In which case the sign corresponds to power consumed by the load the sign corresponds to power s...

Page 9: ...he frequency measurement is performed on the voltage signal 1 Perform a short pressing on the Hz key the display shows the frequency of the measured signal 2 Pressing again returns to the previously d...

Page 10: ...essing unlocks the display 3 12 2 Preselection of MIN MAX mode A short pressing on the HOLD key then on the MIN MAX key preselects the MIN MAX mode A second pressing on the HOLD key makes the MIN MAX...

Page 11: ...12 8 Suppression of automatic stop Hold the HOLD key pressed and bring the rotary switch from the OFF position to the position The unit emits a double beep then the symbol flashes The selected configu...

Page 12: ...x 10 s measurements 6 Battery charge indicator Flashing charge life 1 h Fixed Change battery Automatic OFF after 10 minutes with no action on the switch or the keys reactivation by switching from the...

Page 13: ...V V I I P P C C L L A A M M P P M M I I N N I I 4 6 Variations in operating range 11...

Page 14: ...t the top of the battery cover rear of the clamp and push the battery cover upwards Replace the used battery by a 9 V battery type LF22 observe the polarities Install the battery in its housing then i...

Page 15: ...V V I I P P C C L L A A M M P P M M I I N N I I APPENDIX APPENDIX 13...

Reviews: