Humidity Cabinet
39-1300, 39-1601
4
9907X0093 Issue 3
1.6
Humidity
A wide range of fixed relative humidities can be maintained by adding a suitable
saturated salt solution to the base of the chamber. See Section 4 for details.
2
Installation
2.1
Connect the 3-core cable to a 240V AC, 50Hz mains supply as follows:
BROWN
-
LIVE
BLUE
-
NEUTRAL
GREEN/YELLOW
-
EARTH
2.2
The incubator is internally protected by a fuse, which isolates the cabinet from the
electrical supply if excessive current is drawn. The fuse holder is located on the rear
of the cabinet. The instruments on the control panel will illuminate when the supply is
connected correctly and POWER switch is on.
2.3
General Precautions
A qualified electrician or other competent person must carry out
any electrical work required to install the incubator.
Do not place the incubator in direct sunlight or near a heat
source.
Make sure the incubator is not standing on its electrical supply
cable.
Before the incubator is switched on for the first time, it should
stand still for 2 hours after final positioning to allow the
refrigerant to settle.
Keep all ventilation grilles clear. Failure to do so may result in
component failure.
The refrigeration system gives off heat when operating. Always
ensure there is adequate ventilation around the incubator.
NEVER PLACE ANYTHING ON TOP OF THE INCUBATOR.
Before any cleaning or maintenance work is carried out, the
mains supply must be switched off and the plug removed from
the electrical socket.
3
Temperature Setting Procedure
Switch both the POWER and COOLING switches ON. DO NOT switch the HIGH
HUMIDITY switch on yet!
Please see Section 4.1 for more details.