Portable Conductivity/TDS Meter
513-160
25
9907X0094 Issue 2
5.4.3 Clearing the memory
1) Press and hold
RCL
to recall the stored values from memory.
2) Press
RCL
until
ALL
appears on the display.
3) Press
Read
to delete all measurement results.
CLr
starts blinking on the display.
4) Press
Read
to confirm the deletion
- or -
Press
Exit
to cancel the deletion.
5.5 Self-diagnosis
1) Switch the meter on.
2) Press
Read
and
Cal
simultaneously until the meter displays the full
screen.
Each icon blinks one after the other whereby you can check if all
icons are correctly shown on the display.
After that,
starts to blink and 5 hardkey-icons are shown on
the display.
3) Press any hardkey.
The specific icon disappears from the display.
4) Press each hardkey once.
When the self-diagnosis is completed successfully,
PAS
appears. If the self-diagnosis has failed,
Err 2
appears.
Note
You must press all hardkeys within 1 minute, otherwise
FAL
appears and
the self-diagnosis has to be redone.