Portable Conductivity/TDS Meter
513-160
27
9907X0094 Issue 2
Furthermore, the combined temperature sensor makes the probe automatic
temperature compensating.
Order No: 513-160/10
6.1 For optimum performance, use the following operating procedures:-
6.1.1 Remove all packaging material from the sensor before use.
6.1.2 To prevent carry-over from solution to solution, rinse the sensor with
distilled water between measurements.
6.1.3 When measuring. Make sure the solution is above the cell rings.
6.1.4 Ensure the cell chamber is free of bubbles when measuring.
6.1.5 Allow sufficient time for the sensor to stabilize when measuring
samples at different temperatures. Manual endpoint is advised.
6.1.6 The sensor is not recommended for low conductivity solutions
(˂10 µS/cm).
6.1.7 Clean the probe with distilled water after use.
6.1.8 For calibration and measurement procedures, refer to the instruction
manual.
6.2 Precautions and limitations
6.2.1 Always disconnect the sensor from the meter before cleaning.
6.2.2 Do not expose the sensor to organic solvents, either when cleaning or
when taking measurements.
6.2.3 Do not use the probe outside the recommended temperature range.
6.2.4 Specification and performance of this probe is guaranteed only when
used with this conductivity meter.