Dissolved Oxygen Meter
521-050
12 9907X0046 Issue 1
FAULT
PROBABLE CAUSE
ACTION
Unable to adjust
manual
temperature
Incorrect mode of
operation
Reselect mode
Unable to
calibrate -
Sensor tip may be
tarnished (Ag or Au)
Clean the tip, replace
membrane module
Erratic or slow
response
Air bubbles inside the
membrane cap
Refill or replace membrane
Membrane damaged
Replace membrane
Coating of membrane
due to use in oils or
slurries, strong
solvents, acids or
alkalis
Replace membrane
DO
2
probe defective
Replace DO
2
probe
Inaccurate
calibration -
Zero solution will
absorb oxygen if left
Use fresh zero solution and
recalibrate
Erratic or slow
response
open
Measurement
Incorrect salinity
Reset salinity values
errors
Calibration error
Recalibrate instrument
Contaminated solution
Replace solution
Incorrect standards
Replace standards
Incorrect barometric
pressure
Reset barometric pressure
Unable to
calibrate to 100%
Calibrating to
previously set value
Manually reset value to
100%
Permanently
Membrane not fitted
Fit membrane
reading zero
Membrane damaged
Replace membrane
Membrane not filled
with KCI
Fill or replace membrane
DO
2
probe defective
Replace DO
2
probe
If a fault is identified during the start up self check routine, an error
code will be displayed. This indicates an internal hardware or software
problem. In the event of this, please contact the ELE Service
Department.