Dissolved Oxygen Meter
521-050
5 9907X0046 Issue 1
5
Controls
STO RCL
STO stores the current data reading.
RCL enters recall mode to view the stored readings.
I:O
Switches the instrument on and off. This key should
be pressed and held for 1-2 seconds to operate.
The instrument will automatically switch off after 30
minutes if no key is pressed.
Back light. Pressing this key will illuminate the back
light for 10 seconds. it should be noted that, if used
excessively, this will degrade battery life.
▲ ▼
Enables manual setting of DO
2
calibration values
when performing a calibration, and adjustment of
the manual temperature compensation value,
pressure and salinity.
CAL
This key is used to select and perform DO
2
calibration.
MODE
This key changes the primary measurement
parameter from %DO
2
to mg/l, as required. When
temperature is displayed, holding this key for 3
seconds changes the display to read in °C or °F.
6
Preparation for Use
NOTE: Prior to using the chemicals listed, please refer to the Health &
Safety information at the rear of the manual.
The probe is delivered with the membrane module detached. Prior to use the
membrane module must be fitted to the probe body. A protective sheath is
also supplied. This should be fitted to the probe when not in use. The
sponge contained within the sheath should be wetted with deionised water
prior to fitting. This covers the membrane and keeps it wet, thus preventing
the electrolyte filling from drying out.
a.
Remove the protective sheath from the probe.
b.
Take a membrane module and hold in the vertical position. Fill with
the electrolyte solution (5% KCI) supplied with the instrument. Ensure
no air bubbles are trapped in the bottom of the membrane cap.
Summary of Contents for 521-050