A ball seated screw fitting located on the crosshead accepts ELE load rings and load
transducers up to and including 50 kN.
A socket at the rear of the machine provides a connection for the Electronic Control
and read-out Unit (DSU) for use when transducers to record load and displacement
are required.
3
Specification
Dimensions (approximate)
550 x 400 x 1230 mm (length x width x height)
Maximum vertical clearance 795 mm (platen down, crosshead up)
Minimum vertical clearance
200 mm (platen up, crosshead down)
Horizontal clearance
265 mm
Platen travel (total)
under power
by hand (rapid adjustment)
100 mm
30 mm
70 mm maximum
Speed range
1.27 mm/min
Weight
57 kg
Voltage
10%
Frequency
50 or 60Hz, model dependent
Rated Power
250VA
4
Installation
4.1
Mechanical
DANGER: The CBR 50 is very heavy; it weighs more than 60 kg (132 lbs.). Do
not attempt to unpack, carry or move without proper equipment and sufficient
people to do so safely. Remember, always lift with your legs, not with your
back. If you have a history of back or cardiovascular problems, do not attempt
to unpack or lift the CBR 50.
The machine should be installed on a level bench capable of supporting CBR 50’s
gross weight, and with sufficient space for safe operation of the equipment. The
machine is provided with adjustable feet to compensate for any out of level, or
unevenness, of the bench surface.
4.2
Power Supply
Electrical safety
Warning:
Before removing any covers or performing maintenance repair and service,
isolate from electrical supply by removing mains plug. Where mains supply is
required during these activities, only competent persons should perform the work.
Warning:
Only use the power cable provided or an alternative that is suitably rated
for voltage, current and environment.
Check that the power supply is compatible with the requirements stated on the label
and connect in accordance with IEE regulations or to local requirements.
The power cable is coded as follows: -
Brown wire
L
Live or Power
Blue wire
N
Neutral
Green/Yellow wire
E
Earth or Ground
Note:
exercise extreme caution when using the machine with wet hands.