SoilTest PRO Auto Soils Consolidator (ASC)
25-0500
Original Instructions
13
9901X0275 Issue 2
4.5
Installation Checks
Once the machine is in place and the transducers have been mounted, connection can be
made to the rear panel. The load cell must connect to the lower DIN connector (item 4)
sensor input 1. The LVDT (displacement transducer) must connect to the next DIN connector
(item 5) sensor input 2. Any further displacement sensors associated with manual
consolidation arms connect to the manual sensor inputs (item 6) sensor inputs 3 and 4.
The AC power connector is an IEC C14 fused/filtered appliance switch (item 7). The supply
voltage should be 100 to 240 V AC at 50/60 Hz. The dual pole switch contains two fuses
rated at 1A T. The AC mains supply must be connected to the AC input (item 7) using one of
the IEC 60320 type C13 cable assemblies provided.