SoilTest PRO Auto Soils Consolidator (ASC)
25-0500
Original Instructions
21
9901X0275 Issue 2
Miscellaneous Features
Real-time Clock
A real-time clock is powered by a small LiMn Dioxide battery when no external power is
present. This will maintain the internal clock for up to 5 years even when there is no system
power present and for much longer when the unit is normally powered up
.
Configuration Data
All calibration and configuration data is held in non-volatile memory, and cannot be lost under
normal conditions. All changes to the configuration are stored as they are made.
Error Conditions and Messages
The ASC continually checks critical performance parameters and displays an error message
if a malfunction is detected. If an error is detected, a dialogue box with the error code will be
displayed. If the error has resulted in the machine being shut down and unable to operate
further, this error must be fixed and cleared before the operation can continue.
See
Appendices for a list of error codes and their meanings. If the error persists, or further
assistance is required, contact the ELE service engineer or consult the Service Manual.
Factory Default Values
The ASC can be returned to factory default conditions via
Calibration
˃
Service Functions
˃
Reload Defaults
. The parameters reset include Time and Date format, Language, Units,
User Passcode, Serial Number, Network Address, Machine Number.
Following a Reload Defaults it will be necessary to re-enter the machine Serial Number to
that shown on the rear panel label,
Calibration
˃
Service Functions
˃
Set Serial Number
,
in the form 1944-1-nnnnn.