3.7 Measurement tolerances
The measurement of the electrical conductivity of water is strongly influenced by temperature.
The meter is calibrated to a water temperature of 25°C.
Since MINIguard does not have automatic temperature compensation (ATC), the actual electrical
conductivity of the water differs at other water temperatures.
The measurement tolerance is ±2% per degree Celsius.
Example
Water temperature: ....................................30°C
Deviation: ..................................................5°C
Deviation factor: ........................................10%
Measured electrical conductivity: ...............10 µS/cm
Actual electrical conductivity: ....................11 µS/cm
3.8 Disposal
MINIguard must be disposed of in accordance with local regulations.
You are welcome to send the electronics back to us for recycling.