27
Aut
oSense & Aut
oSense V
oice
• Electromagnetic interference and electrostatic discharge may
interfere with the meter. Do not use it near highly radiant devices,
e.g., mobile phones, microwave ovens.
• Do not perform the test with a damaged test strip.
• Ensure that the test strip is inserted properly. Improper positioning
of the test strip will lead to an inaccurate test result.
• Do not move or remove test strip during measurement.
• Do not perform the test in extreme conditions. External temperatures
that differ from the specified operating temperature range or too
high humidity (e.g., a humid bathroom) might affect the accuracy of
the test result!
• If the meter has been stored in a place that is too cold or too warm,
wait at least 30 minutes before use to ensure the meter and strip
have reached room temperature.
Flagging and marking test results
At the end of each measurement, special flags and marks can be set to
add information to each measurement. The SET button can define 11
different flag and marking combinations. By pushing the SET button, the
following combinations can be set one in sequence: Mark, Premeal,