The black side with the manufacturer’s logo should
face upwards.
Shake the control solution bottle well before use.
Apply a drop of
CareSens
control solution to the
reagent area of the test strip, which is at the tip of
the arrowhead-shaped end of the strip. The meter
will perform the measurement and display the
result in the active unit of measurement (mmol/l or
mg/dl).
Check if the displayed result is within the range of
L1
or
L2
values indicated on the label of the
AutoSense Test
strip vial. If the result is within the
range, your meter is functioning properly.
If the result of the self-test is outside the given value
range:
• Perform self test again.
• Check if the code displayed is the same as the
code on the
AutoSense Test
strip vial.