Chapter 4. Otoacoustic emission (TEOAE and DP-gram)
4.2 Equipment
4.2
Equipment
To make otoacoustic emissions
measurement, you will need the following equipment:
BABYS
C
REEN
unit
OAE
probe
OAE ear tips T04 tree
or
OAE partis Txx (xx size in mm)
•
Connect the
Mini-DIN
plug of the
OAE
Probe
to the
«Audio»
connector of the
BABYS
C
REEN
unit.
4.3
Setup
•
The choice
of
the right EarTip is particularly important for the quality of the measurement.
You can choose from 10 different sizes. The tip must provide the following functionalities:
1.
It must
ensure
that the probe is held securely in the patient's ear.
2.
It
must
not be pressed against a wall of the ear canal.
Check that the 3 small holes at the end of the probe are not blocked. If necessary, replacement tips are deliv-
ered with the unit.
Using an otoscope, make sure that the ear canal is not obstructed by ear-wax.
This operation must be carried out by a qualified person.