9.3. Servicing.
This product should be serviced by a Bibby Scientific Service Engineer or by an agent
on behalf of Bibby Scientific l. If in doubt contact Bibby Scientific. See Section 10.
9.4. Calibration
9.4.1. Recalibration
•
Have a precision reference thermometer of calibrator to hand. Ensure that
OS1=0 and SIM=0
•
Switch the controller off and on again.
•
During the auto-test phase, press buttons button’s
+
and
keep them pressed until the controlled shows 0AD.
•
With buttons
and
selected 0AD or SAD;0AD allows a
calibration of 0, inserting a constant corrective over the while scale of
measurement. SAD allows a calibration of the top part of the measurement
scale with a proportional correction between the calibration point and 0
•
Press
o display the value and then use
+
or
to make the read value coincide with the value measured by the
reference instrument.
•
Exit from calibration by pressing button
.
9.5. Spillage and Decontamination.
In the event of spillage switch off and unplug this product from the mains electrical
supply. Wipe off all excess liquid from the unit and surrounding area using an
absorbent soft cloth.
If in doubt please consult Customer Support. Refer to section 10.
If the equipment has been exposed to contamination, the
Responsible Body is responsible for carrying out appropriate
decontamination. If hazardous material has been spilt on or inside the
equipment, decontamination should only be undertaken under the control of
the Responsible Body with due recognition of possible hazards. Before using
any cleaning or decontamination method, the Responsible Body should check
with the manufacturer the proposed method will not damage the equipment.
Prior to further use, the Responsible Body shall check the electrical safety of
the unit. Only if all safety requirements are met can the unit be used again. The
above procedure is intended as a guide. Should spillage occur with a toxic or
hazardous fluid then special precautions may be necessary.
Decontamination Certificate.
Note:
In the event of this equipment or any part of the unit becoming damaged, or
requiring service, the item(s) should be returned to the manufacturer for repair
accompanied by a decontamination certificate.
Copies of the Certificate are
available from Distributor/Manufacturer. Appendix A of this instructions book
may be copied and enlarged.
At the end of life, this product must be accompanied by a Decontamination
Certificate. See section 6.3 and 6.4
If in doubt please consult Customer support. Refer to section 10.