Page 8 of 24 M6493 Issue 10.3
5. INSTALLATION.
5.1. Electrical safety and Installation.
5.2. This equipment is designed for safe operation under the following conditions:-
•
Indoor use.
•
Altitude up to 2000 meters.
•
Temperatures between -10
°
C and +50
°
C.
•
Maximum relative humidity 80% for temperatures up to 31
°
C decreasing
linearly to 50% relative humidity at 40
°
C.
•
Mains supply voltage fluctuations up to
±
10% of the nominal voltage.
•
Transient overvoltages typically present on the mains supply. (Overvoltage
category II).
•
Applicable rated pollution degree 2.
5.3.
This equipment must be earthed / grounded to a fixed earth / grounded mains
socket outlet. The mains supply is to be earthed / grounded in accordance with
current legislation.
5.4. Ensure only the correct rated mains input fuses are fitted. (Where applicable ensure
the correct mains cable fuse if fitted). See Technical Specification, Section 8 of this
Instruction book.
5.5. Check the voltage on the product data label of this unit. Ensure the rating conforms
to your local supply. Only connect to resistive loads = 1500Watts @ 230V~AC or
750Watts @ 115V~AC. Refer to Technical Specification, Section 8 of this Instruction
book.
5.6. This product must be connected to a mains supply source which incorporates an
RCD or GFCI device.
5.7. The unit is supplied with a moulded mains cord and plug set wired as follows:-
Green /
Yellow
or
Green
=
Earth / Ground
Blue
or
White
=
Neutral
Brown
or
Black
=
Live / line hot.
5.8. If the controller is to be used on a retort rod, make sure it isn’t positioned over any
heating apparatus.
5.9.
If using the wall mounting bracket provided, carefully mark and drill two
suitable holes using the bracket as a template. Ensure there are now cables etc
buried in the wall. Plug the holes with raw plugs and fasten the bracket to the wall in
the correct orientation using the screws provided. Place the controller over the
bracket.