Page 10 of 20 M6847 issue 1.1
7. PRODUCT OPERATION.
7.1. The Paraffin Bath has been designed for easy operation. The illustration below
shows a detailed layout of the product.
1
Bath tank
2
Mains power on neon (White).
3
Heater on neon. (Amber).
4
Temperature control regulation knob
5
Read instruction book label
6
Mains IEC plug socket and protection fuses
7
Data plate label
8
Earth bonding test point.
9
Earth bonding point attention label.
7.2. With the unit disconnected from the mains electricity supply, fill the bath tank with 2
Litres of distilled water.
Note:
For best operation of this equipment, fill the unit with sterilized distilled water.
The acceptable resistivity range is 50K to 1 Megohms (conductivity 1 to 20 micro
Siemens). The acceptable pH range is 7 to 9.
Caution:
Using chlorinated tap water or additives that contain chorine will void the
manufacture warranty. Similary, high purity (deionized) water that does not fall within
a restivity range of 50K to 1 Megohms and pH range of 7 to 9 will void the
manufacture warranty. If you have any questions, please contact Customer support
(see page 15 of this instruction book).
7.3. Plug the unit into the mains electrical supply. Observe the illumination of the White
neon indicating the presence of power.
7.4. The control knob has a temperature scale from 30 to 70 °C.
Set the control knob to the required temperature