12
1. Select a person with normal body temperature as the temperature calibration model.
2. The model shall stand in front of the temperature measuring line, show him/herself in
the center of the infrared display area.
3. Right click to pop up the setting window and select the “IR Position Correction” option.
4. According to the face frame in the visible light image area, draw a frame of the same
position and size in the temperature measurement area. Click the "save" option after the
frame is completed.
3.4 Black body position setting(Factory setting completed)
Set the options to select the black body position setting, and adjust according to the
thermal imaging screen prompts.
Currently, the manual mode is supported to adjust the black body. A person needs to
stand at a distance of 1 to 1.5 meters from the camera to calibrate the black body position.
After the calibration is completed, a "calibration successful" will pop up.
3.5 Black body parameter setting (factory setting
completed)
Currently, the manual mode is supported to adjust the black body, and a person needs to
stand at a distance of 1 to 1.5 meters from the camera to calibrate the body position.