Shake control solution bottle well before use. Drop
CareSens
control solution to the reagent area of test strip, which is at the
tip of the arrowhead-shaped end of the test strip. Meter will
perform measurement and display the result in the active unit
(mmol/l or mg/dl).
Check if displayed result is within the range of L1 or L2 value
indicated on the label of
SensoCard Test
strip's vial. If the
result is within the range, your meter works perfectly.
47
SensoCard