Meter is damaged. Check the meter with
Check-strip
, see “
SensoCard system
checking
” chapter.
7. E-6
error message
Batteries are low. Replace the batteries.
High intensity electrostatic discharge may influence the meter. Please remove the
batteries and keep the
OK
button pressed about 15-20 seconds. Replace the batteries
and turn on the meter. If
E-6
appears again, please change the batteries.
8. E-7
error message
Insufficient amount of blood was applied on strip. Repeat test with bigger drop of
blood.
High intensity electrostatic discharge may influence the meter. Please repeat the
measurement.
9. E-8
error message
Test wasn't performed properly. Repeat test with new test strip and be more careful
when performing each step.