10. E-9
error message
External temperature is too high or low (out of the range 10-40 °C). Wait at least 30
minutes before using the device, until it completely reaches room temperature.
SensoCard
is able to check external temperature. If it detects values not conforming to
its operational requirements, the following error messages will be displayed:
An arrow icon directed downward and the
E-9
error message together indicate low
temperature (below 10 °C).
An arrow icon directed upward and the
E-9
error message together indicate high
temperature (above 40 °C).
11. E-C
error message
This error message may appear during using Code-card. There might be one of the following
problems:
Code-card is damaged. Please check the integrity of Code-card. If you can find any
mechanical damage on Code-card, contact your local distributor.
Code-card insertion or removal was not continuous. Insert and remove Code-card
with a bit quicker expressed move.
52
SensoCard