CALIBRATION
Refer to the METER OPERATION section for test lead connections and measurement procedure.
A/D CONVERTER CALIBRATION
Turn the range selector switch to the 20V position and connect the test leads. Using another meter of known accuracy,
measure a DC voltage of less than 20 volts (such as a 9V battery). Calibrate the kit meter by measuring the same voltage
and adjusting R16 until the kit meter reads the same as the accurate meter (do not use the kit meter to measure its own
battery). When the two meters agree, the kit meter is calibrated. Turn the knob to the OFF position and remove the voltage
source.
SHUNT WIRE CALIBRATION
To calibrate the shunt wire, you will need a 5 amp current source such as a
5V power supply and a 1 ohm, 25 watt resistor. If a 5 amp source is not
available, you can use a lower current (2 amps). If no supply is available, it is
not important to do this test. Set the range switch to the 10A position and
connect the test leads as shown in Figure H. If the meter reads higher than
5A, resolder the shunt wire so that there is less wire between the 10A DC and
COM sockets.
If the meter reads low, resolder the shunt wire so that there is more wire
between the sockets.
1) ACV Range:
750
0 0.0
200
0 0 0
2) DCA,10A Ranges:
200
µ
0 0.0
2000
µ
0 0 0
20m
0.0 0
200m
0 0.0
10A
0.0 0
3) Ohms, Diode and h
FE
Ranges: B indicates blank.
h
FE
0 0 0
Diode ( )
1 B B B
200
1 B B.B
2000
1 B B B
20k
1 B.B B
200k
1 B B.B
2000k
1 B B B
4) DCV Range:
200m
0 0.0
2000m
0 0 0
20
0.0 0
200
0 0.0
1000
0 0 0
-8-
TESTING, CALIBRATION, AND TROUBLESHOOTING
TESTING OF LCD
With no test leads connected to the meter, move the selector switch around the dial. You should obtain the following
readings. A (–) sign may also be present or blinking.
If any of these tests fail:
a) Check that the battery is good.
b) Check the values of resistors R14, R15, R19, R20, R23 - R25.
c)
Check the values of capacitors C1 - C6.
d) Check the PC board for solder bridges and bad solder connections.
e) Check that the slide contactors are seated correctly.
f ) Check that the LCD and zebras are seated correctly.
If the calibration fails:
a) Check the PC board for solder bridges and bad solder connections.
b) Check the value of resistors R7 - R9, R23, and capacitor C3.
10A DC
V
Ω
mA
COM
Figure H
1
Ω
25 Watts
Power Supply
5VDC
–
+
123