CENTRA-R 60/120 US Quick Start Manual
ELGA
Page 4
CENTRA-R 60/120 US MANU40589 Version 02 - 02/16
2. INSTALLATION
2.1 Unpacking CENTRA-R 60/120 US
The following items are supplied:
CENTRA-R 60/120 US.
Electrical supply lead.
Quick reference guide fitted on inside of door.
Consumer Starter Kit 1 or 2.
Installation kit.
Maintenance
Access is required to the front of the unit for regular maintenance procedures. Occasional
access is required to the top of the unit to add chemical during sanitization.
Electrical Supply
For safety reasons it is recommended that this unit be connected to a suitable power outlet
which is easily accessible and local to the unit.
Site
The unit can be floor or bench mounted. Ensure the bench is designed to take a load in
excess of 150kg and that the weight of other equipment is also taken into consideration.
Water Supply
Local potable supply with a means of isolation.
2.2 External services and connections
Connection
Termination
Specification
Unit of Measure
Metric
US / Imperial
Feed water inlet*
½”BSP female thread Flow
>9 l/min
>2.4 USG/min
Pressure
2 to 6 bar
30 to 88 psi
Temperature
4 to 40
C
39.2 to 104
F
Pressurised drain (RO)
½”BSP female thread Maximum flow
10 l/min
2.7 USG/min
Maximum distance to drain
5 m
16 ft
Maximum back pressure
0.2 bar
3 psi
Pressurised drain (loop)
½”BSP female thread Maximum flow
10 l/min
2.7 USG/min
Maximum distance to drain
5 m
16 ft
Maximum back pressure
0.2 bar
3 psi
Application loop feed
½”BSP female thread
Maximum pipe length to
application
25 m
33 ft
Application loop return
½”BSP female thread
Maximum pipe length from
application
25 m
33 ft
Electrical supply
IEC 950
Electrical rating
230 V, 50 Hz ac 13 amp
115 V, 60 Hz ac
* Refer to Operator Manual for feed water specification. Suitability of water supply must be confirmed before
commissioning is commenced.