43
SECTION 5
DIGITAL DATA MANAGEMENT
5.8 Checksums
The checksum for each result line is calculated from the values from the columns BatchID, SampleID, DateTime,
Measurement, Unit, and Device number. A possible result line is as follows:
<Measurement>
<SampleId>01</SampleId>
<DateTime>2016-12-22T12:55:30</DateTime>
<Value>21</Value>
<Unit>mmol/l</Unit>
<DeviceNo>
3400140127</DeviceNo>
<CheckSum>a173e3f754290e77f1b5ac689d3d2486</CheckSum>
</Measurement>
Or:
;
“01”; 2016
-12-
22T12:55:30; 21; “mmol/l”; “3400140127”; “a173e3f754290e77f1b5ac689d3d2486”;
And the above contents are strung together to form:
012016-12-22T12:55:30mmol/l
3400140127
The MD5 checksum for this string is:
A1 73 e3 f7 54 29 0e 77 f1 b5 ac 68 9d 3d 24 86