9MA10011
15/52
4.2 Plant
4.2
PLANT
Name of Unit:
DATA LOGGER
REC. period
min : 000
EN12830:
<=24h / 30s
From this menu you can set:
• Datalogger name ‘Unit Name’:
• Record period ‘REC. period’
- 000 indicates a 30-second record period.
- 001 indicates a 1-minute record period.
- 002 indicates a 2-minute record period, and so on.
On changing the recording period, in real time the line below shows the
degree of conformity with UNI EN 12830 (see First Switch On).
4.3 Not used
4.4 Onboard resources
4.4
ONBOARD RESOURCES
4.4.0
General Info
4.4.1
Analogue Inputs
4.4.2
Analogue Alarms
4.4.3
Digital Inputs
4.4.4.
Digital Alarms
4.4.5
Print config.
(*)
Models with no printer Note that menu 4.4.5 is not available
From this menu, you can
• To set the number and features of Analogue and/or Digital Inputs.
• Print Analogue and Digital Input configuration (models with Printer only, see
Annexe A).
4.4.0 General Info
4.4.0
GENERAL INFO
Number of of probes used:
8
Number of Digitals used
2