When consecutive grouping is enabled, the number of consecutive events is listed in parentheses
along with the grouped item.
11.4 Timing Measurements
The
Overview
windows include timestamps associated with each event captured. Timestamp format is shown as
S
mmm uuu nnn.
Any
Overview
can be linked and unlinked to the
Instant Timing
pane. By default, each
Overview
is linked, and events selected in an
Overview
are displayed in the
Instant Timing
pane. Timing
measurements are easily made with cursors available in the
Instant Timing
pane.
To set a time reference:
1. Right-click in the desired
Overview
.
2. Click on Set Time Reference.
The selected event is assigned a timestamp of zero. Events occurring after the zero-stamped event are incremented
with positive timestamps, and events occurring before the zero-stamped event are incremented with negative
timestamps.
To reset a time reference to the original value:
1. Right-click in the desired
Overview
.
2. Select Reset Time Reference.
The events captured are reset with the timestamp values originally when the capture was taken.
Summary of Contents for USB Explorer 350
Page 1: ...Ellisys USB Explorer 350 User Manual Version 0 83 October 30 2015...
Page 36: ...36 of 157 Exerciser User Interface User Guide Figure 13 Source Capabilities Interface...
Page 41: ...Exerciser User Interface 41 of 157 Ellisys USB Explorer 350 Figure 16 Edit VDM Modes...
Page 61: ...Examiner User Interface 61 of 157 Ellisys USB Explorer 350 Figure 31 Examiner Settings Tab...
Page 86: ...86 of 157 Managing Analyzer Recordings User Guide...
Page 138: ...138 of 157 Analyzer Protocol Error Verifications User Guide...
Page 157: ......