Care and maintenance tasks
72
BA_Elmasolvex_VA_EN_Vers.05.2018
© Elma Schmidbauer GmbH
9.2
Continuous maintenance actions
9.2.1
Change of cleaning and rinsing media
After view (visual inspection of the cleaning and rinsing media
in the opened media tanks) and/or in the event of diminishing
cleaning result.
There are several possibilities available for compliance with the
intervals for the media change:
If a service life of the media has been saved using the display,
an appropriate message is shown on the display.
If the media in the media tanks seem to be increasingly
contaminated or the cleaning parts after cleaning no longer
appear to be clean, the media must be changed.
Only permitted media (cleaning / rinsing media) are permitted to
be used for safety reasons and to prevent cleaning machine
damage.
Observe the instructions for recommended media and
limitations for unsuitable / not permissible media (
Chapter 8
)
.
Fire and explosion hazard!
Observe the applicable safety regulations for handling
solvents.
Keep all kinds of ignition sources away.
Prevent ignition sparks from electrostatic discharge. Discharge
possible electrostatic charges (body charge) by touching any
grounded equipment before handling flammable media: e.g.
water tap, metal surface of the case of the cleaning machine or
use ESD protective equipment (ESD arm band).
Remove the relevant media tank from the cleaning machine.
The media tank must not be removed during running cleaning
programs.
Drain the relevant media tank and clean if necessary.
Proceed as described in
Chapter 4.4
for refilling the media
tanks.
Check the correct closing function of the quick couplings
(
Chapter 4.4.1
) before reconnecting the media tanks.
Used media must be disposed of in accordance with the
regulations. No disposal via the sewer system! Dispose of the
used media in accordance with the national disposal regulations
for the media (see safety data sheet).
Recommended
interval
Inspection criteria
Reminder on the
display
Manual monitoring
Only use permitted
operating materials
DANGER
Procedure
ATTENTION
Disposal of used
media