Product description
4.7
Description of operating elements
Illustration 4.7 view operating element Denta Pro H
Temperature display
shows the actual temperature of the
cleaning bath.
Temperature range 20°C – 35°C: green LEDs
Temperature range 40°C:
yellow LED
Temperature range 45°C – 70°C: red LEDs
Flashing LED warn that the temperature of the cleaning bath
has exceeded the temperature set for the selected cleaning
programme. This excess temperature protection ensures that
the temperature of the cleaning bath will not be too high for the
selected cleaning process: excess temperature protection for
the programmes Instruments and Drills /rotating instruments
from 42 °C, for the programme Dentures from 60°C.
Individual operation
mode
Start key
to start the cleaning programme or to start permanent
operation.
The green LED flashes in stand-by mode and is lighted in
operating mode.
Extra key
special operating mode for heavy contamination or
for a large number of cleaning items.
For the cleaning without cleaning programme
at a
temperature of 50°C or 70°C (Denta Pro H only).
Preset cleaning programmes
provide specially defined
combinations of different cleaning modes: degas – boost –
normal – sweep, the various modes can be distinguished
through different operating sounds. Each programme operates
at a specific predefined temperature (Denta Pro H only).
A
B
C
D
E
F
A
B
D
C
E
F
© Elma Schmidbauer GmbH