How to open the unit
24
SM_Elmasonic_S_02.2018_EN
© Elma Schmidbauer GmbH
Illustration 6.3. Fastening of the bottom plate
6.4
Elmasonic S 60, S 100 – S 900 H
The bottom plate is screwed to the unit through the unit feet. In
addition, the bottom plate is screwed to the cooling body of the
PCB control with 2 screws.
1. Pull the mains plug!
2. Place the unit onto the workplace upside down.
3. Unscrew the 4 unit feet.
4. Unscrew the two visible recessed head screws.
5. Carefully lever the bottom plate loose with a screw driver or
a similar tool
illustration 6.1.
).
Ilustration 6.
3
. Unscrew the two recessed head screws and the 4 unit
feet
Fastening of bottom
plate
How to proceed
2x
4x
B
B
B
B
C
A