How to replace the heating
© Elma Schmidbauer GmbH
SM_Elmasonic_S_02.2018_EN
55
13.1
Heating wiring S 450 H and S 900 H
On the units S 450 H and S 900 H the heating is switched via a
relay. The zero wire is connected with the strip terminal directly
from the interference filter. The phase is conducted to the relay
through the PCB control.
Illustration 13.5. Relay and strip terminal for the wiring of the heating.
Terminal distributor heating to heating wiring
Relay
Supply cable voltage supply / phase (black) from interference
filter
Connection phase switched relay >> terminal distributor heating
(black)
Supply cable control relay, phase switched from PCB control
(grey)
Supply cable mains voltage / zero to relay coil supply
Supply cable mains voltage / zero (white) from interference filter
S 900 H: Connection mains voltage (zero) to terminal distributor
of heating on the other side of the transducer tank
S 900 H: Connection mains voltage (phase) switched to
terminal distributor of the heating on the other side of the
transducer tank