Putting into operation / Trial run
64
SM_Elmasonic_S_02.2018_EN
© Elma Schmidbauer GmbH
18
Putting into operation / Trial run
For checking and trial operation of the unit after repair please
observe the instructions given in this section. We recommend a
trial run over several hours (approx. 3 - 4 h).
Place the unit on a dry and solid workplace. Ensure that the
workplace is sufficiently ventilated!
Risk of electrocution due to entering humidity!
Keep the unit dry.
The unit inside is splash-proof. To prevent electrical accidents
and damage keep the workplace and the unit housing dry.
•
Allowed ambient temperature during operation:
+5°C to +40°C
•
Allowed relative humidity of air during operation: 80 %
•
No condensation allowed
The unit must be connected to a shockproof socket only.
18.1
How to fill cleaning liquid
Fill the cleaning tank of the unit with a sufficient quantity of a
suitable liquid before switch-on.
Dry running of the transducer tank can damage the unit.
Ensure that the cleaning tank is correctly filled during operation
(marking on the tank wall).
Make sure that the selected cleaning chemical is perfectly
suitable for the use in an ultrasonic bath and follow the
instructions on dosage and material compatibility.
In general, flammable products are not allowed for use in an
ultrasonic bath. Observe the safety warnings in
section 17.2.
Limitations concerning other cleaning media are also listed in
the warnings in
section 17.2.
Fresh cleaning liquid is saturated with air which impairs the
cleaning effect of the ultrasonic activity. Sounding the liquid
over a period of several minutes before the actual cleaning
process is started will eliminate the microscopic air bubbles
from the liquid.
CAUTION
Ambient conditions
Mains connections
Observe filling level
NOTE
Suitable cleaning
media
Prohibited cleaning
media
Degassing of the
liquid