Trouble shooting
© Elma Schmidbauer GmbH
BA_Elmasonic S_EN_Vers. 05.2019
35
Fault
Possible cause
Remedy
Unit does not heat up;
LED temperature dark
turning knob temperature in
„0“ position
switch on turning knob
temperature
unit is switched off
switch on unit with key on/off
fault of electronics
return unit to supplier or
manufacturer
No heating function;
LEDs of LED temperature
blink alternately ("running
light") = fault indication
heating
fault of electronics
switch unit off and on
if fault is indicated again:
return unit to supplier or
manufacturer
Unsatisfactory heating-up
period
loss of heating energy
use cover (optional
accessory equipment)
no mixing of cleaning liquid
e.g. switch on ultrasound
(see section 7.2)
Unit produces boiling
noise during heating-up
no mixing of cleaning liquid
e.g. switch on ultrasound
(see section 7.2)
Set temperature is
exceeded
temperature sensor does not
measure the average
temperature (no revolution)
mix liquid manually or by
means of ultrasound
set temperature too low,
ultrasonic energy heats up the
liquid more than required
(physical process)
for low set temperatures do
not switch on heating
switch on ultrasound for
short periods only
No operational functions;
LEDs of LED ultrasound
and LED temperature blink
alternately ("running light")
= fault indication
programme control
fault of electronics
switch unit off and on
if fault is again indicated:
return unit to supplier or
manufacturer