Important safety warnings
© Elma Schmidbauer GmbH
BA_Elmasonic S_EN_Vers. 05.2019
7
2.3
Intended use as medical device
Ultrasonic pre-cleaning of
- surgical and medical instruments.
Only reusable medical devices that are approved for ultrasonic
cleaning and authorised for the reprocessing are permitted to
be cleaned (see information of the medical device manufacturer
according to, e.g. EN ISO 17664).
The pre-cleaning of medical devices with Elmasonic S-Line
products does not replace the subsequent cleaning, disinfection
or sterilisation in automated standard processes (e.g. cleaning
and disinfection unit (CDU) or autoclave).
The user is responsible for checking the cleaning result.
Ultrasonic cleaning machines are exclusively intended for
ultrasonic irradiation of objects and liquids. Thereby, no
flammable liquids are permitted to be used directly in the
cleaning tanks.
The machine is only permitted to be operated by trained
personnel and not by children. Operation and placement must
be performed in accordance with the conditions and media
defined in the operating manual.
The service intervals and regional regulations for checking the
equipment must be complied with.
2.4
Safety instructions on the machine
Observe operating instructions!
Observe warnings and safety instructions given in the operating
manual!
This symbol warns about the risk of injury from hot surfaces and
liquids.
The unit cannot be disposed with household waste! Observe
regional waste regulations!
Intended purpose
Intended use