How to replace the PCB interference filter
© Elma Schmidbauer GmbH
SM_Elmasonic_S_02.2018_EN
33
9
How to replace the PCB interference filter
9.1
Elmasonic S 10 (H) – S 40 (H)
The PCB interference filter is screwed to the back of the unit.
The mains input socket is soldered to the PCB interference filter
and must be replaced with the PCB.
9.1.1
Electric connections
The units manufactured in 10.2005 or before are fitted with the
PCB interference filter illustrated below.
When ordered as a spare part the new version will be delivered
(
see description for year of manufacturing 11.2005 or later
). For
mounting follow the instructions on mounting the new version.
Illustration 9.1. Connections PCB interference filter
Connections for heating element
(mains voltage) on units
with heating. Polarization of the cables is irrelevant.
Connections to PCB control
(mains voltage). Polarization of
the cables is irrelevant.
Version 10.2005 and
before
A
B
A
B