Initial operation
© Elma GmbH & Co KG
BA/X-tra/0908
9
Dry running of the transducer tank can cause damage to the
unit.
Ensure that the cleaning tank is filled to the optimum level
during operation (see tank marking).
ATTENTION
Ensure that the chosen cleaning chemical is suitable for
treatment in an ultrasonic bath and observe the instructions on
dosage and the compatibility of the material.
Suitable cleaning
agents
We recommend the use of the cleaning agents listed in section
6.1.
Flammable products are not allowed for use in an ultrasonic
bath. Observe the instructions given in section 6 (cleaning
media).
Prohibited cleaning
agents
Restrictions concerning further cleaning media are also
indicated in the instructions given in section 6.
For queries please contact the manufacturer or your
supplier.
Freshly mixed cleaning liquids are saturated with air which
lessens the cleaning effect of the ultrasonic activity. By
sounding the liquid over a period of several minutes before the
cleaning process the tiny air bubbles in the liquid are
eliminated.
Degassing of liquids