Trouble shooting / List of malfunctions
© Elma GmbH & Co KG
SM/Transsonic_LC/1104/D
13
6.2 T420(H);
LC20H
Malfunction
1. Check:
pilot lamp
ultra-
sound
2. Check:
heating
switch
- if lighted
3. Check:
fuse
printed
board
Possible causes
Remedy
Check the unit for possible damages and traces of liquid inside (e.g. leak in the
tank, liquid/humidity entering from outside)
No ultrasound /
weak ultrasound
(Heating OK)
OK
OK
OK
•
output
transformer
•
output
capacitor
•
scorching in
transducer
system
•
replace
component
•
replace
complete printed
board
•
open/check
transducer
system
No ultrasound
(Heating OK)
OFF
OFF
faulty
•
faulty printed
board (BZW
diode;
transistors;
diodes)
•
short circuit in
heating
disconnect
heating cable;
check again)
•
replace
component
•
replace
complete printed
board
•
replace heating
No ultrasound
(Heating OK)
OFF
OK
OK
•
interference
filter coil
•
timer
•
replace
component
•
replace
complete
interference
filter
No ultrasound
(Heating OK)
OFF
OFF
OK
•
interference
filter
•
mains cable
•
replace
complete
interference
filter
•
replace mains
cable
No heating;
No ultrasound
OFF
OFF
OK
•
scorched
interference
filter
•
faulty mains
cable
•
replace
complete printed
board
•
replace mains
cable
Summary of Contents for LC130H