background image

 

 

1. Press 

+

 and 

-

 at the same time to enter the menu. 

2. Step to 

INTERNAL

 with 

M

.

  

3.  Set the number of samples (1 - 36) per hour with 

+

 and 

-

.  

4. Confirm with 

=

.

 

5.  Choose between FLon and FLoF. Use

 

+

.

 

6.  Confirm and start sampling with 

=

.

 

7.  When FLon is chosen the display shows "IntF" until the level sensor is activated.

 

4.1.2 External control 

 
External controlled sampling is performed after pre-setting of the parameters that effects the sampling 
frequency. 
 
External control device is connected through the Amphenol-contact at the lower side of the sampler.  

 
A. Flow control sampling 

 
Samples are taken in proportion to the flow; the sampler gets a signal from a flow meter. 
 

1. Press 

+

 and 

-

 at the same time to enter the menu 

2. Step to 

MAX FLOW

 with 

M

.  

3.  Set the numeric value of the actual max flow with 

+

 and 

-

 keys, keep the key pressed to accelerate 

the numeric updating.  

4. Confirm with 

=

 

(the display flashes).  

5.  Stop the flashing with 

+

 or 

-

. Put the decimal point at the correct place with 

+

 (at the left) and

 

-

 

(at the right). 

6. Confirm with 

=

.

 

7.  The sampler steps automatically to 

INTERVAL

.

 

8.  Set the numeric value of the interval, number of m³ between samples, with 

+

 and 

-

 keys keep the 

key pressed to accelerate the numeric updating  

9. Confirm with 

=

 

(the display flashes).  

10. Stop the flashing with 

+

 or 

-

. Put the decimal point at the correct place with 

+

 (at the left) and

 

-

 

(at the right). 

11. Confirm with 

=

.

 

 

NOTE! Control that the interval < max flow, and that max flow / interval not exceeds 36 samples/h. 

 

12. The sampler steps automatically to 

MAN SAMPLE

.

 

13. Step to 

ANALOG

 with 

M

14. Set the signal range (0 - 20 mA alt. 4 - 20 mA) with 

+

 key. 

15. Confirm and start sampling with 

=

 

(the display shows the actual flow in % of max flow).

 

 
Example of settings at flow controlled sampling  

Ex.1 

Max. flow 

5,5 m

3

/h 

Interval 2 

m

3

 between samples 

 
The sampler takes samples according to 5,5/2 = 2,75 samples/h, one sample every 22nd minute at max. flow. 

 
Ex.2 

Max. flow 

3,6 m

3

/h 

Number of samples  3 samples/h 
 

Summary of Contents for VA/V3

Page 1: ...Water Sampler VA V3 v 2 6...

Page 2: ...ECTION 5 3 3 5 3 3 1 INTERNAL 5 3 3 2 PULSE 5 3 3 3 ANALOG 6 3 3 4 MAX FLOW 6 3 3 5 INTERVAL 6 3 3 6 MAN SAMPLE 6 3 3 7 TOTAL SAMPLES 6 3 3 8 ALARM 6 3 4 KEYS 7 4 SETTINGS 7 4 1 SETTING OF SAMPLING CO...

Page 3: ...e front panel The LCD display gives you running information during both programming mode and sampling sequences 1 1 2 Sampling unit The sampling unit consists of a sample bottle in and outlet tubes an...

Page 4: ...rackets at the rear side of the sampler NOTE Avoid placing the sampler in direct sunlight 2 3 ELECTRICAL INSTALLATION External control For connection to external control a 4 wire cable with Amphenol c...

Page 5: ...2 3 3 Connection of a potential free contact level sensor etc Connection is made through the black and black wire on the external control cable 2 4 CHECKING THE CONNECTIONS Before the main supply is t...

Page 6: ...hen water is present in the vacuum tube the sampler immediately turns off all active functions and empties the sample bottle Error code E 08 is shown on the display In this case contact Elmacron AB 3...

Page 7: ...he operators command 3 3 7 TOTAL SAMPLES Every correct sample without error is accumulated 3 3 8 ALARM When an error appears the red diode at ALARM lights up At serious errors the display flashes and...

Page 8: ...le is taken at the operators command 1 Press and at the same time to enter the menu 2 Step to MAN SAMPLE with M 3 Start a sampling sequence with B Time related sampling Flow on Flow off function When...

Page 9: ...the display flashes 5 Stop the flashing with or Put the decimal point at the correct place with at the left and at the right 6 Confirm with 7 The sampler steps automatically to INTERVAL 8 Set the num...

Page 10: ...ling with 4 2 ALARM EMPTY SAMPLE BOTTLE 1 Press and at the same time to enter the menu 2 Step to ALARM with M 3 Choose between ALon alarm on and ALoF alarm off use key 4 Confirm and start sampling wit...

Page 11: ...on 5 1 SAMPLING SEQUENCE During sampling sequence you can follow the steps in the sampling sequence at the display Display 1 The discharge valve closes 2 Suction tube purge 3 Sample in 4 Discharge of...

Page 12: ...n the screw at the bottle holders upper part 4 Release the bottle pull it straight up 5 If the bottle holder needs to be cleaned loosen the screw at the bottle holders lower part and pull the holder s...

Page 13: ...vacuum tube NOVA if it is OK CH06 Test of menu diodes All diodes on the menu line lights up 6 3 CALIBRATION The water sampler is always pre calibrated when delivered Pre calibration is only necessary...

Page 14: ...l in the tank Check tubes and connections E 03 Sequence collision to many pulses Reduce the number of pulses E 04 The pulse length is to long 5 s Check pulse transmitter E 05 The pulse length is to sh...

Page 15: ...Number of samples 1 36 samples h Main supply 220 V Protection IP65 Max suction height 6 m Alarm output Potential free contact max load 24VDC Control Internal time External potential free pulse Externa...