Page 3 of 4 

V1/July 2015

 

 

 
Mother-Baby Local Pairing 

When bracelets with mother infant pairing capabilities (P/N 5-WTD41120-0) are used, then the pairing with an Elpas Charm (P/N 5-

BTE00433-1) may be performed as follows: 

1. 

Make sure the Mother Bracelet is within a 15cm Range of the charm. 

2. 

Connect the Charm band.  

3. 

When the Charm red LED flashes, press the button of the Mother Bracelet and hold it until the Charm green LED flashes. The LF 

IDs of both charms are paired. 

4. 

For a match test, press the button of the Mother Bracelet and hold it until the Charm green LED flashes. 

Note:

 To pair the The mother Bracelet and Infant Charm through EV2, see Eiris Configuration Guide, section 9.3.2. 

 Note:

 In case of a duplicate ID Trouble Alert, submit a support ticket with Elpas at 

www.elpas.com

 

 

Note:

 To configure an Elpas Badge Alert for the Infant Protection Charm, see Eiris Configuration Guide, section 5.15 

False Tamper Alerts 

If you use a Security Band and it produces false Tamper alerts, clean the band thoroughly as dirt can act as a buffer between the 

contacts. If false Tamper alerts persist, replace the band. 

Cleaning & Disinfection Procedures 

Use an appropriate antibacterial disinfectant such as Dispatch® Hospital Cleaner Disinfectant with Bleach from Caltech Industries, Inc 

(

http://www.caltechind.com

) to clean the Bracelet. 

Since ‘Cleaning Procedures’ may vary according to facility guidelines, thus the procedures given below are for illustrative purposes only: 

Option 1 

– Using Dispatch Disinfectant Spray 

1. 

Lightly wet a disposable towel with Dispatch spray 

2. 

Do not saturate the towel 

3. 

Wipe the outer surfaces of the bracelet 

4. 

Next wipe the bracelet with a dry disposable towel 

5. 

Allow the bracelet to air dry 

6. 

Return the clean bracelet to inventory or usage 

7. 

Dispose of used towels per facility policies 

Option 2 

– Using Dispatch Disinfectant Towels 

1. 

Open a new Dispatch pre-moistened towel 

2. 

Wipe the outer surfaces of the bracelet 

3. 

Next wipe the bracelet with a dry disposable towel 

4. 

Allow the bracelet to air dry 

5. 

Return the clean bracelet to inventory or usage 

6. 

Dispose of used towels per facility policies 

 