Alarm Conditions
In the section "Alarm Mode" the possibility to select the inspection range has been added. The following
options are available:
All data:
all recorded values are taken into consideration when assessing the alarm status
(based on the configured alarm conditions)
Last "alarming ON" period only:
only the measured values between the last pressing of the
"Alarming ON" button and the last pressing of the "Alarming OFF" button on the device are
considered in the assessment. If the "Alarming ON" button was pressed last, i.e. the alarming is
still active, all measured values since that time are taken into account.
All "alarming ON" periods cumulative:
all measured values recorded in phases with activated
alarming are considered in the assessment.
Alarming ON/OFF
Only if one of the last two options are selected, the alarming can be activated ("Alarming ON") or
deactivated ("Alarming OFF") during measurement mode.