6. Threshold value and alarm function

Sensor selection

The ECOLOG-NET LP4 datalogger has a feature for monitoring threshold values. Threshold values and 
alarm parameters are defined in the "Setup of Alarm Parameters" window. Threshold values are only mon-
itored using sensors that are selected in the "Datalogger setup - sensors" window. 

5.1 

Datalogger 

setup

6.1 Conditions

Threshold values
Alarm Delay Time
Alarm output

- The measured value must be outside the defined tolerance range, i.e. the measured value is higher 

than the maximum allowable threshold value or lower than the minimum allowable threshold value.

- The threshold violation must exceed the defined alarm delay times.
- The threshold violation remains active until the measured value reaches the defined tolerance range 

again. 

No alarm conditions are:

- Cable break of the sensor cable
- Short circuit of the sensor cable
- Readings outside of the measuring range

 

10.1 

Datalogger display and elproLOG ANALYZE

Summary of Contents for ECOLOG-NET LP4

Page 1: ...ECOLOG NET LP4 EL6006Ee ECOLOG NET LP4 Operation Manual...

Page 2: ...r normal conditions All main items of the software product comply with the operating instructions and the information in the help file The guarantee does not apply to consumables disposable batteries...

Page 3: ...ng mode 8 2 8 Overlaying graphical data 8 3 Configuring the datalogger 9 3 1 Assign network address 9 3 1 1 Desktop installation 9 3 1 2 Documentation 9 3 2 Installing the datalogger 9 3 2 1 Installat...

Page 4: ...10 Status and error messages 24 10 1 Datalogger display and elproLOG ANALYZE 24 10 2 Datalogger status in elproLOG ANALYZE 24 11 Maintenance 25 12 Technical data 26 12 1 Measuring ranges and Accuracy...

Page 5: ...ct to CE marking The manufacturer guarantees that this product complies with the relevant guidelines see 13 Declarations of conformity This product must be disposed of in accordance with WEEE Waste el...

Page 6: ...ft side of the housing and on the front a display and a membrane keyboard 1 Network connection 2 Connection for 4 Pt100 sensor 200 C 600 C 3 USB connection 4 Power supply contact inputs and alarm outp...

Page 7: ...anteed that the loggers will function properly if the datalogger is exposed to tempera tures which exceed the specified threshold range Experience has shown that the battery freezes at approx 50 C tha...

Page 8: ...y The datenlogger ECOLOG NET has a power save mode which switches off the display As a result mea surements are only made during the defined log interval Four small circles located in the display indi...

Page 9: ...igned to the datalogger during this workstep This work should be per formed at the place of use prior to final installation e g in an office 3 1 2 Documentation Documentation for the implemented confi...

Page 10: ...g the datalogger 3 2 5 Verification of the installation Check the installation and ensure that sensor positions alarm parameters and network addresses are cor rect Test connectors can be used as an ai...

Page 11: ...rm or when the self sustaining function is selected and an alarm is registered 6 2 Window Setup of Alarm parameters Jump to menu item 5 Reset alarm display and alarm contacts as well as entry of the r...

Page 12: ...ly Definition of the alarm thresholds 6 Threshold value and alarm function Reset alarm This function is used to acknowledge an alarm mes sage Set date and time Used to adjust the integrated datalogger...

Page 13: ...Datalogger setup 6 1 Conditions Threshold values Alarm Delay Time Alarm output The measured value must be outside the defined tolerance range i e the measured value is higher than the maximum allowab...

Page 14: ...nction Alarm thresholds Input fields for the lower and upper threshold values Hysteresis The Hysteresis is used to avoid fluttering of the alarm contact A change of state in the contact only occurs wh...

Page 15: ...semiconductor switch break contact to datalogger GND Semiconductor contact alarm 2 Only reacts when there is an alarm This is a semiconductor switch make contact to datalogger GND Internal Buzzer Ser...

Page 16: ...1 cycle time 2 Alarm is logged ECOLOG NET Alarm contact responds immediately ANALYZE 2 Alarm measurement ANALYZE Alarm protocol with 2 entries alarm on off MONITOR Represents an alarm for 7 cycle time...

Page 17: ...alarm and therefore it is not recognized as an alarm 5 Reset of delay time by short threshold value violations ECOLOG NET Alarm contact does not react Even in the case of repeated short threshold valu...

Page 18: ...of the contacts is checked during the update of the monitor data Status changes between two updates are not detected Both inputs are busy with dual functions In each case only one function should be...

Page 19: ...ction 8 1 Pt100 sensors 1 4 Pt100 Connector metal lized DB9 female Part no 800504 Schema Pt100 3 wires Pt100 4 wires cable length 12m Part no ECA_PT100_SILxx Pt100 4 wires cable length 12 30m Part no...

Page 20: ...20 EN ECOLOG NET LP4 EL6006Ee Pin assignments and connections Pt100 4 wires cable length 30m Part no 800520 CABLE LENGTH 12 1 Measuring ranges and Accuracy...

Page 21: ...logger operating volt age 6 sc Alarm contact relay contact 7 sa Alarm contact relay contact 8 Gnd Ground 9 Gnd Ground Alarm contact relay contact Relay contact represented in dead state Connection sa...

Page 22: ...und 5 V in External power supply logger operating volt age 6 Gnd Ground 7 Gnd Ground 8 Gnd Ground 9 Gnd Ground FOLLOW THE SAFETY AND APPLICATION INSTRUCTIONS OF THE POWER SUPPLY UNIT Contact input Ala...

Page 23: ...or VDC 500mA Alarm Threshold value violation or cable break Alarm Contact open Datalogger does not require external power supply Semi conductor contact Use only for DC voltage DCV Switch load max 24VD...

Page 24: ...of the batteries is displayed in the elproLOG ANALYZE status information Battery replacement 11 Maintenance Firmware The version of the firmware is documented in the status Display elproLOG ANALYZE CA...

Page 25: ...ium 3 6V 2100mAh AM3 LR6 AA Battery An energy consumption counter is used to monitor datalogger battery life For this reason only the specific manufacturer recommended battery should be used Do not re...

Page 26: ...1 C 0 2 C 400 1 C 500 0 C 0 1 C 0 3 C 500 1 C 600 0 C 0 1 C 0 5 C This information applies to the ECOLOG NET LP4 sensors at room temperature and without Pt100 Length Cable typical errors 50m 4x0 25mm2...

Page 27: ...EN 27 ECOLOG NET LP4 EL6006Ee Technical data IT ES EN EN FR IT ES 12 3 Dimensional view 1 Battery for a battery exchange the rear wall has to be removed 2 Screws for the attachment of the rear wall...

Page 28: ...ension cable for Pt100 sensor 0 5m for cable guides e g refrigerators ECA_4C_xxM_M12 Connection cable Pt100 up to 30m with M12 plug ECA_LIYCY4X0 34 4 wire connection cable with M12 plug for cable leng...

Page 29: ...EN 29 ECOLOG NET LP4 EL6006Ee Declarations of conformity IT ES EN EN FR IT ES 13 Declarations of conformity...

Page 30: ...er 04 03 2011 First edition replaces EL6003Bk A Gubler 21 12 2011 a Minor changes A Gubler 01 10 2012 b New Graphic A Gubler 03 06 2013 c No marking of the sensor cable required A Gubler 13 01 2014 d...

Page 31: ...ation 9 E ECOLOG NET LP4 6 elproLOG ANALYZE 9 24 elproLOG CONFIG 9 elproLOG MONITOR 9 Environmental conditions 7 Error messages 11 24 Explosion protection 6 Extended setup 12 Extended setup 12 F Firmw...

Page 32: ...Operation Manual ECOLOG NET LP4 EL6006Ee 11 2016 Head Office ELPRO BUCHS AG Lang ulistrasse 45 9470 Buchs SWITZERLAND e mail swiss elpro com For local representatives see www elpro com...

Reviews: