Data Logger DL210
122
Elster-Instromet GmbH
B-3 Encoder, pulse and message inputs
2 signal inputs with common ground for reed contacts or transistor switches . No
extraneous voltages or currents should be fed in.
Designation
I1... I2
Cable connection
Screw terminals; 0.5 ... 1.5 mm² (solid);
With flexible cable use wire-end sleeves.
Special features
each input can be parameterised and sealed separately,
Input I2 can only be used as a signalling input
Max. cable length
max. 30 m depending on ambient conditions (EMC)
Nominal data
a) Use of input I1 as LF pulse or signalling input
and / or input I2 as signalling input
Open-circuit voltage
U
0
»
3 V
Internal resistance
R
i
»
1 M
W
Short-circuit current
I
s
»
3 µA
Switching level "on"
R
e
< 1 M
W
Switching level "off"
R
a
> 1 M
W
Pulse duration
t
e
³
50 ms
Space duration
t
a
³
50 ms
Counting frequency
f
£
10 Hz
b) Use of input I1 as encoder input with Namur encoder
Open-circuit voltage
U
0
»
8.2 V (7...9V)
Internal resistance
R
i
»
1 K
W
Short-circuit current
I
s
»
8 mA (7...9mA)
Switching level "on"
I
e
£
1.2 mA
Switching level "off"
I
a
³
2.1 mA
Baud rate
2400 Bd
Data format
7 data bits, even parity bit, 1 stop bit ("7e1")
c) Use of input I1 as encoder input with SCR encoder
Open-circuit voltage
U
0
»
6 V (4.75...6.5V)
Internal resistance
R
i
»
1 k
W
Short-circuit current
I
s
»
6 mA (4.75...6.5mA)
Switching level "on"
I
e
> 3 mA
Switching level "off"
I
a
< 3 mA
Baud rate
300 Bd
Data format
7 data bits, even parity bit, 1 stop bit ("7e1")
Summary of Contents for DL210
Page 2: ...Data Logger DL210 2 Elster Instromet GmbH ...
Page 9: ...Data Logger DL210 Elster Instromet GmbH 9 ...
Page 10: ...Data Logger DL210 10 Elster Instromet GmbH ...
Page 11: ...Data Logger DL210 Elster Instromet GmbH 11 Part 1 Device description relevant to calibration ...
Page 101: ...Data Logger DL210 Elster Instromet GmbH 101 Part 2 Description of the Initial operation ...
Page 118: ...Data Logger DL210 118 Elster Instromet GmbH Appendix ...