50
Assembly
, Connection and Putting into Operation
Once you have changed all of the digits, press the ENTER key in
order to confirm your entry. The input value can be deleted by
using the ESC key.
5.3.1.5 Configuration of the outputs
The cp values (pulse constants) for the outputs can also
be configured using the enSuite software.
The following functions are preset ex-factory for the outputs:
Output
Function
DA1
Pulse output for volume "V1" (cp.A1= 0.1)
DA2
Pulse output for volume "V2" (cp.A2= 0.1)
The cp value of an output indicates how many pulses are emitted
per cubic meter. A cp value of 0.1/m3 (0.1 pulses per m3) means
that e.g. a pulse is emitted per 10m3.
To change the cp values for output 1 or output 2, move the cursor to
the "Serv." register and go to "cp.A1" (for output 1) or "cp.A2" (for
output 2) via the following path:
Serv.
Outputs
Output 1
cp.O1
or
Serv.
Outputs
Output 2
cp.O2
Press the ENTER button
a digit of the cp value will flash.
Move the cursor to the digits using the
or
arrow keys and
update them using the
or
arrow keys.
Press ENTER, so that the value setting is confirmed. The input
value can be deleted by using the ESC key.
Aside from the settings described here, a range of other
functions can be configured for the outputs, e.g. status
information or time-synchronous pulses.
A complete description can be found in the DL230
Application Manual that can be downloaded under
www.elster-instromet.com.
5.3.1.6 Setting the daylight saving time
Move the cursor to the "Serv." register and to the "MdTim" value
(daylight saving mode) via the following path:
Serv.
Date and time
MdTim
Summary of Contents for DL230
Page 1: ...Operating manual Data Logger DL230 ...
Page 8: ......
Page 105: ...Appendix 105 English manual 73 022 387 ...
Page 106: ...106 Appendix 9 2 EC Declaration of Conformity ...
Page 107: ...Appendix 107 9 3 ATEX Type Examination Certificate 9 3 1 Associated apparatus for Zone 0 1 ...
Page 108: ...108 Appendix ...
Page 109: ...Appendix 109 ...
Page 110: ...110 Appendix ...